SPECIFICATIONS

PERFORMANCE CHARACTERISTICS

Power Supply:

2 x 1.5V. alkaline batteries (AAA)

Battery Life:

Approximately one year when testing up to four times a day.

Result Range:

0–600 mg/dL (0–33.3 mmol/L). Values of 600 mg/dL

 

(33.3 mmol/L) and above displayed as HI.

Display Type:

Liquid crystal.

Blood Source:

Whole blood (capillary or venous). Do not use samples

 

that contain fluoride (grey top tubes).

Haematocrit Range:

25–60%.

Dimensions:

Length, 10.9 cm. (4.3 inches); Width, 6.6 cm. (2.6 inches);

 

Height, 3.0 cm. (1.2 inches).

Weight:

127.5 grams (4.5 ounces) with batteries.

Operating Temperatures:

15°–35°C (59°-95°F).

Operating Humidity Range:

0–90% relative humidity (non-condensing).

Altitude:

One TouchBrand Glucose Monitoring Systems, based on One

 

Touch technology, have been shown to give equivalent perfor-

 

mance when tested at altitudes up to 2200 meters (7200 feet).

Memory:

Up to 250 records with time and date.

Data Port:

RS-232 serial communications with special LifeScan cable for

 

interface with LifeScan InTouchDiabetes Management System

 

Software.