MESSAGE | PROBLEM |
NOT OK | 1. If appearing when you turn |
| your meter on, your meter may |
| have an electronic problem. |
| 2. If appearing at the end of a |
| test, your meter may have been |
| moved while applying blood or |
| during the test process. |
WHAT TO DO
1.Call your authorized LifeScan representative.
2.Repeat the test, try not to move the meter.
| 3. If appearing during the test, | 3. Repeat the test with the new |
| the strip has been removed at | test strip. |
| the WAIT PLEASE screen. |
|
|
|
|
58 MG/DL or | The check strip test result is | Clean the meter and check strip if |
3,2 MMOL/L | outside the acceptable range. | necessary. |
(for example) |
| Repeat the test. |
✓NOT OK |
|
|
7:40 AM |
|
|
|
|
60