Chapter4 HardwareOverview
NI5102 User Manual 4-14 ni.com
TriggerSources

The Scan Clock, Start Trigger, and Stop Trigger signals can be generated

through software or supplied externally as digital triggers or as analog

triggerso n one oft hei nput channels or the TRIG channel. Figure 4-10

shows the different trigger sources. In addition, Scan Clock is available

from a source (counter) internal to the NI 5102.

Figure4-10. ScanClock, Start Trigger, and Stop TriggerSignal Sources

CH0
Programmable
GainAmplfier
Programmable
GainAmplifier High
Level COMP
COMP
Low
Level
+
+
Gain
Gain
CH1
TRIG
Analog
Buffer
ADC0
ADC1
Analog
Mux
Digital
Mux StartTrigger
AnalogTrigger
CircuitOutput
RTSI<0..6>
(WhereAvailable)
PFI1,PFI2
Software
Digital
Mux ScanClock
AnalogTrigger
CircuitOutput
PFI1,PFI2
RTSI<0..6>
(WhereAvailable)
InternalScan
StopTrigger
Digital
Mux
Analog
Trigger
Circuit
AnalogTrigger
CircuitOutput
7
2
2
a.Analog Trigger Circuit
b.Trigger Sources c.Scan Clock Sources
StarTrigger (PXI)