AppendixA Specifications
NI5102 User Manual A-6 ni.com
Maximum Working Voltage(Signal voltagepl us common-mode voltage)
Channelto earth ......................................5 V, Installation Category I
Channelto channel .................................5 V, Installation Category I
EnvironmentalOperatingtemperature ............................0 to 55 °C
Storagetemperature ................................–20 to 70 °C
Relative humidity ...................................10%to 90% noncondensing
Maximum Altitude .................................2000meters
Pollutiondegree ......................................2
Indoor use only
SafetyMeets the requirements of the the followingstandards for safety for
electricalequipment for measurement, control, and laboratory use:
EN 61010-1:1993/A2:1995, IEC 61010-1:1990/A2:1995
UL 3101-1:1993, UL 3111-1:1994, UL 3121:1998
CAN/CSA C22.2 NO. 1010.1:1992/A2:1997
Electromagnetic CompatibilityCE,C -Tick,and FCC Part 15 (Class A) Compliant
Electricalemissions ................................EN 55011 Class A at 10 meters
FCCPart 1 5A above 1 GHz
Electricalimmunity ................................Evaluated to EN
61326:1997/A1:1998, Table 1
Note Forfull EMC comp liance, you must operate this devicewit h shielded cabling. In
addition, all coversand filler panels must be installed. See the Declaration of Conformity
(DoC)for this product for any additional regulatory compliance information. To obtain the
DoCfor this product, click Declaration of Conformity at ni.com/hardref.nsf/.This
website lists the DoCs byp roduct family.Select the appropriate product family, followed