©NationalInstruments Corporation A-1 NI5102 User Manual
A

Specifications

Thisap pendix lists the specifications of the NI 5102. These specifications
are typical at 25 °C unless otherwise stated. The operating temperature
rangeis 0 to 50 °C.

Input Characteristics

Numberof input channels ...................... 2 single-ended,
simultaneously sampled
Inputimpedance ..................................... 1 M±1% in parallel
with 25 pF ±10 pF
(Impedance increases
with attenuating probes)
CH0,CH1, TRIG
ADC resolution...................................... 8 bits, 1 in 256
Maximum sample rate
Internal............................................ 20 MS/s each channel
in realtime mode
Externalsample clock..................... 20 MS/s
Minimum high or low time...... 24 ns
RIS mode ........................................1 GS/s
Minimumsample rate ............................1 kS/s (internal/external)
Maximum input range............................ ±5000 V, DC + peak AC < 5 Mhz
(with a X1000 probe)
±500V, DC + peak AC < 15 Mhz
(with a X100 probe)
±50V,DC+peakAC<15Mhz
(with a X10 probe)
±5 V,D C+ peak AC < 15 Mhz
(with a X1 probe)