VI Basic Operations

MICROSCOPE

 

 

 

 

6 Using the Microscope Control

 

 

Precautions for 90i setup

1Make sure the observation method is set to bright-field microscopy.

2Make sure the objective used is

N.A. 0.04 to 0.95

x1 to x100.

(Phase contrast, liquid immersion, polarization, Hoffman, multi-immersion, and industrial-use objectives may not be used.)

3Make sure the objective information is set at each address on the nosepiece.

Other precautions

1Autofocus relies on contrast at the center of the view field. Thus, the area of the target to

be examined will not come into focus unless it is at the center of the view field. The microscope system may have difficulty focusing accurately on samples like the ones indicated below.

Low contrast samples

Samples densely dyed (dark) over the entire view field

Samples prepared for methods other than bright-field microscopy

Samples that lack light and dark patterns running horizontally along the screen (lateral direction in the diagram shown at right)

Detection range

2Since autofocus responds to the highest contrast part of the sample, the focused position of the sample may not always be the area you want to observe. In such cases, adjust focus visually by turning the focusing handle of your microscope.

3The autofocus may occasionally misfocus due to external disturbances. If the image fails to come into focus, try autofocusing again.

4The microscope system may operate erratically if dirt or dust accumulates in the following areas.

Top or bottom faces of the slide glass

Tip of the objective

Surface of the condenser

In the middle of the optical path

- 145 -