| .TOOL. | VI Basic Operations |
|
|
|
3Measuring
8)Press to return to the previous menu. Any calibration values will be erased unless they have been saved.
NOTE: Do not save the calibration value when “E” is checked in the calibration screen. Otherwise, an incorrect calibration value will be saved.
< When using the magnification of the objective and relay lenses >1) | Select a number in which you want to save the calibration value from |
| s1 | through | s7 | . | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
| Drag the mouse pointer over one of numbers | s1 | through | s7 | , and the calibration | |||||||||||||||
| information for that number will be displayed in the Info area for your confirmation. | |||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| NOTE: If the microscope connected to your system is the Nikon 80i/90i microscope | |||||||||||||||||||
|
| with motorized or intelligent nosepiece, note that each calibration value call | ||||||||||||||||||
|
| number is linked to the hole number of the nosepiece. This means you | ||||||||||||||||||
|
| cannot change calibration values by selecting |
|
| through |
|
| . The value | ||||||||||||
|
| s1 |
| s7 | ||||||||||||||||
|
|
|
|
|
|
|
|
|
| |||||||||||
|
| is selected automatically from | s1 | through |
| s7 |
| based on the hole | ||||||||||||
|
| number of the nosepiece you are currently using. |
|
|
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
| |||||||||||
|
|
|
|
|
|
|
|
|
| |||||||||||
2) | Press | Calib. | to bring up the calibration screen shown below. |
|
|
|
|
|
|
3) Click the numeric value input box to display the numeric keypad. Use the keypad to enter the magnification of the objective.
4) Click the numeric value input box to display the numeric keypad. Use the keypad to enter the magnification of the relay lens.
5)If you want to attach information to the calibration value you've created, click the Info input box and enter text for the information you want to attach. This information will prove helpful because you can confirm Info when you call up the calibration value at a later time.
6) | Check the Set number to confirm that the calibration number is correct, then click | Save | ||
|
|
|
| |
| to save it. | Save | is disabled if no value is entered. |
7)Press to return to the previous menu. Any calibration values not saved are deleted. NOTE:
*If the microscope connected to your system is the Nikon 80i/90i microscope with motorized or intelligent nosepiece and the calibration method you selected is "using the magnification of the objective and relay lenses," you do not need to enter the magnification of the objective. If the camera head is connected to the rear port of the DIH, you do not need to enter the magnification of the relay lens. If you calibrated with one objective, other connected objectives will be also calibrated automatically with their magnifications being converted.
*If you selected the "using the magnification of the objective and relay lenses" calibration method, keep in mind that the calibration value may incorporate a certain amount of error attributable to optical errors of the objective. With the Nikon 80i/90i microscope connected, errors may range up to 10%.