Hardware 3-7

Steps for RS-232/422 Configuration

Standard Units, Only Early Production Units, Only

1. Turn the ChartScan system’s power “OFF.” 1. Turn the ChartScan system’s power “OFF.”

2. Ensure the COMM SELECT micro-switch (on the

rear-panel DIP switch) is in the “0” (down)

position.

2. Ensure the COMM SELECT micro-switch (on the

rear-panel DIP switch) is in the “1” (up) position.

3. Adjust micro-switches 2 through 9 (for parameter

preferences), according to the following table for

standard units.

3. Adjust micro-switches 2 through 8 (for parameter

preferences), according to the following table for

early production units.

Note: Early production ChartScan units do not have a Calibration Enable/Disable Pushbutton. If you have an

early production unit, refer to the right-hand DIP switch illustration, above.

RS-232/422 Serial Settings (Standard Units) RS-232/422 Serial Settings (Early Production Units)

Selection Micro-
Switch
#
Setting Selection Micro-
Switch
#
Setting
COMM
SELECT
1 0 – Serial Communication; RS-232 or
RS-422 depending on switch 9.
1- Selects IEEE 488.
COMM
SELECT
1 0- selects IEEE 488 interface card.
1- Selects RS-232/422.
HANDSHAKE
(H/S)
2,3 0, 0- No Handshake
0, 1- Software Handshake Only1
(XON/XOFF)
1, 0- Hardware Handshake Only
(RTS/CTS)
1, 1- Both Hardware and Software
Handshake
HANDSHAKE
(H/S)
2,3 0, 0- No Handshake
0, 1- Software Handshake Only1
(XON/XOFF)
1, 0- Hardware Handshake Only
(RTS/CTS)
1, 1- Both Hardware and Software
Handshake
PARITY 4,5 0, 0 - No Parity
0, 1- Odd Parity
1, 0- Even Parity
1, 1- N/A
PARITY 4,5 0, 0 - No Parity
0, 1- Odd Parity
1, 0- Even Parity
1, 1- N/A
SERIAL
BAUD RATE26,7,8 0,0,0- 300 Baud
0,0,1- 600 Baud
0,1,0- 1200 Baud
0,1,1- 2400 Baud
1,0,0- 4800 Baud
1,0,1- 9600 Baud3
1,1,0- 19200 Baud3
1,1,1- N/A
SERIAL
BAUD RATE26,7,8 0,0,0- 300 Baud
0,0,1- 600 Baud
0,1,0- 1200 Baud
0,1,1- 2400 Baud
1,0,0- 4800 Baud
1,0,1- 9600 Baud3
1,1,0- 19200 Baud2,3
1,1,1- N/A
RS-232 or
RS-422
9 0- Selects RS-232 serial*
1- Selects RS-422 serial*
* (switch #1 must be in the “0” position)
CALIBRATION
MODE
9 0- Calibration Mode Disabled
1- Calibration Mode Enabled
(normal position is “0”)
1 ChartView software will not run when Handshake is set for “Software Handshake Only,” (0, 1).
2 It is recommended that you do not exceed 9600 Baud unless you have firmware version 1.3 or greater. The firmware version can be read in
ChartView’s title bar.
3 At 9600 and 19200 Baud Rates, Hardware Handshake should be set to “Hardware Handshake Only” (1, 0). We recommend that you do not
exceed 9600 Baud unless you have firmware version 1.3 or greater. The firmware version can be read in ChartView’s title bar.