3-18 ChartScan User’s Manual
CSN14/HV/S High Voltage Card with Safety Jack Connectors

The CSN14/HV/S card contains 16 differential input channels and is for use with “safety jack” type connectors. Each

CSN14/HV/S card is capable of measuring analog input signals on any of the following three programmable ranges:

±2.5 volt, ±25 volt and ±250 volt.

CSN14/HV/S, High Voltage Card with Safety Jack Connectors

Number of Inputs:
16 differential; programmable by channel for input range
Input Connectors:
Safety Jack
Voltage Range, Accuracy and Resolution:
± 2.5V; ± 0.02%; 78.14 µV/bit
± 25 V; ± 0.02%; 781.4 µV/bit
± 250 V; ± 0.02%; 7.81 mV/bit
Digital Filtering:
Used for AC line cycle noise rejection.
Selectable off, 1-256 samples in a binary count.
Temperature Coefficient: < 0.01%/°C
Input Impedance: 10 M typical
Input Bias Current: 20 pA maximum.
Maximum Common Mode Voltage:1 500 VDC or 500 VAC peak
Maximum Normal Mode Voltage: 500 VDC or 500 VAC peak
Common Mode Rejection: 100 dB typical.
Maximum Allowable Input: 500 VDC or 500 VAC peak
Channel-to-Channel Isolation:
500 VDC or 500VAC peak
Channel-to-Power Ground Isolation:2
500VDC or 500 VAC peak
Note 1: Specified for coupling impedance >30 M and common mode frequency < 60 Hz. 700 VDC or 700 VAC peak ; 325 VDC
or 325 VAC peak if used in the same system with a MTC/24 scanning module, before equipment damage oc curs.
Note 2: 325 VDC or 325 VAC peak if used in the same system with a MTC/24 scanning module.