Omega Vehicle Security 1400 manual Normal Mode, CH1 and CH2 Assigned to a Chart

Models: 1400

1 273
Download 273 pages 33.71 Kb
Page 94
Image 94

 

 

 

Acquisition Setup

Parameter

 

 

Options

Event Configuration

Trigger: Set a trigger scan in the Acquisition by using one of the following:

 

 

Keyboard - Use Console Trigger dialog box to start trigger scan.

 

 

Note: If the Keyboard Trigger is selected the use of Pre-trigger is disabled.

 

 

External TTL - Use an external signal to start the trigger scan.

 

 

Channel Value - Use specified value of a given channel.

 

 

Alarm - Use an Alarm value to start the trigger scan.

 

 

Absolute Time - Use a time of day to start the trigger scan.

 

 

Note: If Absolute Time is selected, the use of Auto Re-arm is disabled.

 

Stop:

Set a stop scan in the Acquisition by using one of the above selections or Count (as

 

 

specified below). Keyboard isn’t applicable. If a MultiScan, ChartScan, or NetScan is

 

 

connected and High-Speed Single-Channel (Burst Mode) is selected, then “Count from

 

 

trigger” is the only Stop available for the event configuration.

 

Count:

Specify a number of scans to be read before a stop scan is set.

Acquisition Parameters

Scan Counts:

Specify the number of scans in an acquisition

 

 

 

Pre-trigger

 

 

 

Post Stop

 

Average weight: Average weight can be specified only when in the Normal Mode. The value for

 

 

 

the average weight determines the number of samples to be averaged when in

 

 

 

Normal Mode. Values of 1, 2, 4, and 8 are for high-speed multiple channel

 

 

 

applications for thermocouples and DC voltages where noise filtering is not an

 

 

 

issue.

 

 

 

Values of 16, 32, 64, 128, and 256 are recommended for AC noise rejection.

 

Scan Intervals:

Time between scans (frequency of scans)

 

 

 

Pre-trigger and Post Stop

 

 

 

Post-trigger

 

 

 

Use one Interval

High Speed Setup:

Allows for channel and scan rate entries for high-speed, single channel mode.

Note: These parameters cannot be altered while an acquisition is in progress.

Note: If Normal Mode is selected, channels assigned to a chart must be (and will be) enabled in the Acquisition.

The assigned channels will be “ON” in the Channel Configuration portion of the Channel and Alarm Setup spreadsheet and can not be turned off while assigned to a chart. However; additional channels (not assigned to charts) can still be enabled and disabled from the Channel and Alarm Setup sheet.

Example. In the figure to the left:

(a) Channel 1 and Channel 2 can not be turned “OFF.” Each is assigned to a chart.

(b) Both Channel 3 and Channel 4 can be turned “OFF.” Neither is assigned to a chart.

(c) An attempt to remove a chart-assigned channel will result in a message box similar to that in the figure.

Normal Mode, CH1 and CH2 Assigned to a Chart

4-38

ChartView Software Reference

ChartScan User’s Manual

Page 94
Image 94
Omega Vehicle Security 1400 manual Normal Mode, CH1 and CH2 Assigned to a Chart