| 13 | Troubleshooting | |
|
|
|
|
| |
Symptoms |
|
| Checks/Remedies | ||
LED does not irradiate. |
|
| Did you select the target channel for the product type? (When selected, the channel | ||
|
|
| switch is lit or flashing.) |
|
|
|
|
| →Select the target channel when setting the product type. | ||
|
|
| If the target channel is selected but does not irradiate, check if the irradiation time | ||
|
|
| is set to 0.0sec for the channel program for the target product type. | ||
|
|
| →If so, set the irradiation time to 0.0sec or longer. | ||
|
|
|
| ||
When a warning message is displayed: |
|
|
| ||
Temperature warning |
|
| The channel connection indicator changes to orange (mixture of green and red) | ||
|
|
| when the LED head temperature has reached the warning level. If you continue the | ||
|
|
| operation in this condition, the product life will be significantly shortened, and the | ||
|
|
| temperature may exceed the upper limit, causing an emergency stop. | ||
|
|
| →Set the UV intensity level (%) lower or improve the heat radiation performance | ||
|
|
| of the LED head to keep the temperature lower than the warning level. | ||
Time warning |
|
| When the total irradiation time of the LED reaches the point 30 hours before the | ||
|
|
| preset LED replacement point, the time warning is issued and “x100hrs” will be | ||
|
|
| lit. When the time reaches the LED replacement point, the irradiation process will | ||
|
|
| make an emergency stop. |
|
|
|
|
| →Prepare a new LED head. After replacement, clear the total irradiation time. | ||
When an error occurs: |
|
|
|
|
|
Interlock emergency stop |
|
| The emergency stop contact was opened during irradiation. | ||
“The interlock contact | was | opened | →Check the emergency stop contact or the external input terminal block. | ||
during irradiation.” |
|
| Close the contact and then restart irradiation. |
|
|
|
|
| * If the emergency stop contact is open, normal resetting is not available. | ||
|
|
| Hold down SET to forcibly reset the interlock. |
|
|
|
|
|
| ||
Connection error |
|
| It is possible that the connection cable or LED head was broken. | ||
“The LED was disconnected during | →Reset the error status, check the connection of the cable and LED head with the | ||||
irradiation.” |
|
| channel in question, and replace them if necessary. | ||
|
|
|
| ||
LED wire broken |
|
| The current detection was stopped during irradiation. It is possible that the | ||
“LED circuit opening was detected | connection cable or LED head was broken. |
|
| ||
during irradiation.” |
|
| →Reset the error status, check the connection of the connection cable and LED | ||
|
|
| head with the channel in question, and replace them if necessary. | ||
|
|
|
| ||
LED |
|
| An overcurrent was detected during irradiation. It is possible that the connection | ||
“LED | detected | cable or LED head was |
|
| |
during irradiation.” |
|
| →Reset the error status, check the connection of the cable and LED head with the | ||
|
|
| channel in question, and replace them if necessary. | ||
|
|
|
|
|
|
LED temperature error |
|
| The LED temperature exceeded the upper limit. |
|
|
“The LED temperature | reached an | →Set the UV intensity level (%) lower or improve the heat radiation performance | |||
abnormal temperature.” |
|
| of the LED head to keep the temperature lower than the warning level. | ||
|
|
|
| ||
LED time error |
|
| The total irradiation time of the LED reached the preset LED replacement point. | ||
“The total irradiation time reached the | (Or, it is possible that the LED replacement point was set ahead the total irradiation | ||||
replacement point.” |
|
| time.) |
|
|
|
|
| →Reset the error status, replace the LED head, and then clear the total irradiation | ||
|
|
| time. (Or, go to the mode setting screen and correctly set the LED replacement | ||
|
|
| point.) |
|
|
LED calibration error |
|
| Calibration was stopped before completion. Or, the calibration was not successful. | ||
“Calibration uncompleted” |
| It is possible that the setting is too high, or there is a problem in the dedicated UV | |||
|
|
| sensor (option). |
|
|
|
|
| →Reset the error status, check the connection of the UV sensor, review the setting, | ||
|
|
| and then retry calibration. |
|
|
|
|
|
|
|
|
57