Storage: 30 days (after opening) at
Barcode labels are provided for use with the diluent. Before use, place an appropriate label on a 16 × 100 mm test tube, so that the barcodes can be read by the
L2HOZ: 3 labels
L2SUBM: Chemiluminescent Substrate
L2PWSM: Probe Wash
L2KPM: Probe Cleaning Kit
LRXT: Reaction Tubes (disposable)
L2ZT: 250 Sample Diluent Test Tubes (16 × 100 mm)
L2ZC: 250 Sample Diluent Tube Caps
CCCM: A
Also required
Distilled or deionized water; test tubes; controls.
Assay Procedure
Note that for optimal performance, it is important to perform all routine maintenance procedures as defined in the IMMULITE 2000 Operator's Manual.
See the IMMULITE 2000 Operator's Manual for: preparation, setup, dilutions, adjustment, assay and quality control procedures.
Recommended Adjustment Interval: 4 weeks.
Quality Control Samples: Use controls or plasma or serum pools with at least two levels (low and high) of homocysteine.
Expected Values
Homocysteine levels can vary with age, gender, geographical area and genetic factors, therefore it is important for laboratories to establish their own reference ranges based on their local populations. Literature suggests a reference range of 5 – 15 µmol/L for adult males and females,11 but also notes that men tend to have higher levels than women, and that postmenopausal women tend to have higher levels than premenopausal.12
One hundred and twenty samples from apparently healthy adult male and female volunteers age
4
using the IMMULITE 2000 Homocysteine procedure. The samples were drawn in heparinized plasma tubes and kept on ice prior to separation of the plasma from the cells. The median value was 7.7 µmol/L, with a central 95% range of
5.0 – 12 µmol/L.
Consider these limits as guidelines only. Each laboratory should establish its own reference ranges.
Limitations
Specimens from patients who are on drug therapy involving
Although analysis of the parent compounds of carbamazepine, phenytoin,
Heterophilic antibodies in human serum can react with the immunoglobulins included in the assay components causing interference with in vitro immunoassays. [See Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988:34:27- 33.] Samples from patients routinely exposed to animals or animal serum products can demonstrate this type of interference potentially causing an anomalous result. These reagents have been formulated to minimize the risk of interference; however, potential interactions between rare sera and test components can occur. For diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings.
Performance Data
See Tables and Graphs for data representative of the assay's performance. Results are expressed in µmol/L. (Unless otherwise noted, all were generated on heparinized plasma samples.)
Calibration Range: 2 to 50 µmol/L.
IMMULITE 2000 Homocysteine