Getting Started
680A03 TekConnectProbe Interface Module Instruction Manual
Toinstall the sampling module into the front panel of the TekConnect Probe
InterfaceModule, perform the following tasks:
CAUTION. Sampling modules areinherently vulnerable to static damage. Always
observe static-safeprocedures and precautions as outlined in your sampling
module user manual.
1. Verifythat the sampling oscilloscope is powered off; if necessary,toggle the
front-panelOn/Standby switch.
2. Insert thesampling module into the front-panel compartment of the
TekConnectProbe Interface Module as shown in Figure 4.
3. Push the module in slowly with firm pressureto seat it.
4. Once the module is seated, turn the hold-down screwto tighten the sampling
module in place. See Figure 4.
5. Torelease the module later, loosen the hold-down screw fully,and then slide
the ejectorlever.
Ejectorlever
Samplingmodule
Hold-downscrew
Figure 4: Installing the samp ling module into the TekConnectProbe Interface
Module
NOTE. Toremove the sampling module, first loosen the hold-down screw; then
use the ejector lever on the interface m oduleto eject the sampling module.
Installing the SamplingModule