Getting Started
80A03 TekConnectProbe Interface Module Instruction Manual 9
Performing a FunctionalCheck
Toverify that the 80A03 TekConnect Probe Interface Module is functioning
properly,perform the following procedure.
Refer to Table2 for a list of recommended equipment.
Table2: Recommended equipment for perfor ming a functional check
Item description Recommended example
Samplingoscilloscope TektronixCSA8000 or TDS8000
Sampling module Tektronix80E02, 80E03, or 80E04
TekConnectprobe Tektronix P7000 Series
Probeti p adapter TektronixP7225, P7240: 015-0678-xx
TektronixP7260: 067-1456-xx
TektronixP7330, P7350: 067-0419-00
Semi-rigid SMA cables (2) Tektronixpart number 174-4857-xx
SMA cable Tektronixpart number 174-1427-00
50 termination, SMA-type Tektronixpart number 015-1022-xx
Torquewrench SMA 5/16, 8 in-lb
Toverify the functionality of Channel A, perform the following tasks:
1. Initializethe system:
a. Power on the sampling oscilloscope.
b. Waitfor the boot up process to complete.
c. Press the front panel DefaultSetup button.
2. Connect the signal source:
a. On the sampling oscilloscope, connect the SMA cable to the Internal
Clock Output. See Figure 7 on page 10 fordetails.
b. Connect the Probe TipAdapter with 50 termination to the other end of
the SMA cable.
3. Set up the trigger system:
a. In the UI applicationtoolbar, select Internal Clock from the Trig List
menu box.