Getting Started
80A03 TekConnectProbe Interface Module Instruction Manual 11
6. Verifythe channel operation:
a. Verifythe LED on the TekConnect Probe Interface Module for the
channel undertest is Green.
b. On the sampling oscilloscope verifythe probe model, serial number,
scale, dynamicrange, impedance, and other information by selecting the
Utilities >System Properties >Probes menu for the channel under test.
NOTE. The sampling oscilloscope displays the first six characters of the
TekConnectprobe model and truncates the remaining characters. For
example, the oscilloscopew ill identify a P7350SMA probeas P7350S.
The impedance value displayedfor P7330, P7350, and P7350SMA
probesis for single-ended configuration. The displayed value is one half
the value listed on the probelabel.
c. On the sampling oscilloscope,verify that the extender cable length is
listed as 1 meterin the Utilities >System Prop erties >Sampling
Modules menu.
d. On the sampling oscilloscope,verify that the signal displayed is a square
wave of approximately1 Vp-p with a 5 s period.
e. Adjust the offsetand verify that the display signal moves vertically to
the offset level.
f. Adjust the vertical scale and verify that the displ ayed signal height
changes.
Toverify functionality of the TekConnect Probe Interface Module Channel B,
performthe following tasks:
1. Connect the probeto C hannelB instead of Channel A.
2. Repeat steps 5 through6 above.
Toverify that the oscilloscope, sampling module, or TekConnect probe is
performingproperly, run the calibration, compensation, or functionalverification
procedurerecommended for the individual product. For detailed instructions,
refer to the user manuals that accompanied t hese products.
Functional Checks forOther Equipment