Getting Started
10 80A03 TekConnectProbe Interface Module Instruction Manual
4. Connect the probe:
a. Attach the P7000 Series probe to the Channel A input of the 80A03
TekConnectProbe Interface Module.
b. Install the probe tip onto the probe tip adapter.
Figure7: Equipment set upf or functionalcheck
5. Set up the channelfor test:
a. On the sampling oscilloscope,press the channel button to select the
channel thatis connected.
The Channel button lights, the sampling module LED lights, and the
selectedchannel on the sampling oscilloscope becomes active.
b. Turnthe Vertical scale knob and set the sensitivity to displaya waveform
similar to the oneshown in F igure7.
c. Turnthe Horizontal scale knob to indicate 2 s/div.