Reference
22 80A03 TekConnectProbe Interface Module Instruction Manual
CAUTION. Toprevent dama ge from electrostatic discharge, install 50
terminations (Tektronixpart number 015-1022-xx) on the sampling-module
connectors beforeremoving the sampling modules from an instrument or when it
is not in use. Storethe sampling module in a static-free container, such as the
shipping container. Whenever you move the sampling module from one instru-
ment to another,use a static-free container to transport the sampling module
Toprevent dam age to the sampling module, discharge to ground anyelectrostat-
ic chargethat may be present on the center and outer conductors of cables
beforeattaching the cable to the sampling module.
Toprevent dam age to the sampling module, do not create an ESD antenna by
leaving cablesattached to the sampling-module input with the other end of the
cable open.
Toprevent dam age to the sampling module or instrument, never install or
removea sampling module when the instrument is powered-on.
Always use a wrist strap (providedwith your instrument) when handling
sampling modules or making signal connections.Wear anti-static clothing and
work in a static-freeworkstation when using sampling modules.