Operations 4

7.After pressing ACK in step 6, the new sensor is flushed

for 10 minutes while the measured O2 level is displayed. The 10-minute flush may be restarted by pressing RE- SET, or aborted by pressing SKIP.

8.When sensor installation or alternate sensor test are complete, the system will return to the main Analyze mode menu.