4 Operations

13. If the O2 cell current is too strong or too weak to span, an error message will appear. This will not occur if the user-entered span gas concentration is correct. Press ACK to retry, or ESC to abort the span.

14. If the span is successful, the system will return the main Analyze mode menu. Check to see that the span value is correct.

4-10

TELEDYNE ANALYTICAL INSTRUMENTS

Page 38
Image 38
Teledyne 3160 manual Operations