3 Installation
1.The sensor compartment is located in the lower
2.Make sure that sample gas is flowing through the analyzer, then shut off the flow completely. This insures that air does not diffuse into the analyzer.
3.Inside the compartment should be two (one) stainless steel cylindrical sensor block(s), each with a clamp lever across the front. The bottom half of the clamp face is bent outward. To release the clamp, pull the clamp lever up and toward you slowly.
4.Releasing the clamp releases the stainless steel block cap on the bottom of the sensor block. Pull the cap downward to remove it. There is a guide pin to the rear of the sensor block that aligns the cap to the block.
Note: It is important to minimize the amount of time that the new cell is ex- posed to air in order to reduce the time required for the reading to drop to zero.
5.Remove the
6.Carefully guide the sensor and cap into the block, putting the guide pin through the hole in the back edge of the cap.
7.While holding the cap in place, squeeze the clamp down firmly until the notches on the clamp lock onto the block side pins.
8.Immediately start the sample flow, and set to about 2 SCFH. The analyzer now needs to be zeroed and calibrated.
3.6Electrical Connections
All of the electrical connections are located on the analyzer back panel. The analyzer is shipped with all of the electrical connections intact, with no assembly or installation required. The power cord receptacle is located in the lower center of the panel. The voltage selection terminal and the fuse block are in the same fitting directly above the power cord receptacle. There are four output signal connectors with screw terminals in the upper