1 Introduction
The 3160 features six
•
•
•
•
•
•
An AutoRanging feature automatically selects the range that is appropriate for a given reading. For example, if the O2 level exceeds 100% of the current range, the analyzer switches to the next higher range. When desired, a manual override feature allows a particular range of interest to be
With the AutoSpan feature, the analyzer automatically calibrates at scheduled intervals, and automatically performs electronic zeroing and sensor settling detection during calibration.
Five
•a personal computer loaded with the Teledyne Remote Analytical Control Software (TRACS)
•any terminal or terminal emulation software
•a custom computer program
Several analyzer functions, including troubleshooting, can be run from your PC through the phone line to the analyzer.
The analyzer has separate sample and span gas ports, which allow the installation of an external source of span gas for calibration without interfering with the sample gas line.
An optional scrubber installed into the sample system allows sample gas to be used as a zero gas after treatment with the scrubber. If the scrubber is installed, pneu- matic valves on either side of the scrubber open automatically during zeroing.
Note: Units equipped with a scrubber should only be used with inert gases and saturated hydrocarbons.
The analyzer output is not affected by minor changes in the flow rate. Analyzer performance is stable under minor mechanical vibration.
TELEDYNE ANALYTICAL INSTRUMENTS |