Operations 4

Cold Start-Up

1.A cold start will automatically occur the first time you start the analyzer. In any other situation, you can cold start by turning the analyzer off and holding down the red key below the LCD screen while you turn the power back on. Continue to hold the red key until the screen comes on. There will be a two minute wait for system initialization before the second screen.

A warm start will take you directly to step 5.

2.Set the time using the SELEC, UP and DOWN keys, and press ACCPT.

3.Select sensor A or B as the active sensor.

(This menu is omitted if only one cell block is installed.)

4.If you have not done so, connect the sample gas to the sample port and press ACK.

5.Using the valve and meter on the sample gas supply, set the flow rate to 2 SCFH and press ACK.

If you have warm-started, this is the second screen you will see.

6.The main Analyze mode menu will appear, and the

measured O2 level of the sample gas will display. The system should first be zeroed, then spanned, to obtain the most accurate O2 measurements.