5 Maintenance & Troubleshooting

Analyzers equipped with a scrubber should only be used with inert gases and saturat- ed hydrocarbons.

5.2Troubleshooting

To troubleshoot the analyzer, identify the problem in the following list and then

follow the procedures to correct the problem. If you cannot identify the problem, or if you cannot resolve the problem after following the corrective procedure, call a service repre- sentative.

Symptom

Cause

Correction

LCD or LED display will not

Faulty LCD or LED.

Return the unit to factory.

light.

 

 

Display too bright.

 

Remove analyzer cover. Adjust

 

 

potentiometer located behind

 

 

LCD display.

 

 

 

No analog output.

Faulty electrical connection.

Remove analyzer cover. Check

 

 

PC board connections and

 

 

make sure that they are firmly

 

 

seated.

 

 

 

Zero gas signal >2 PPM.

Bad sensor.

Replace sensor.

 

High O2 level.

Check zero gas inlet connec-

 

 

tion.

Slow response/recovery time.

Bad sensor.

Replace sensor.

Displays “--” value.

 

Re-zero sensor.

 

 

 

Can't calibrate (with sensor installed). Display reads “cell too strong” or “cell too weak.”

a)The wrong span gas concentration has been entered.

a)Re-enter span gas concen- tration.

b)Replace or reinstall sensor.

No sample gas flow.

a)

The pneumatic air supply

 

 

pressure is too low, making

 

 

the valves stick.

 

b)

The sample supply pressure

 

 

is too low.

a)Raise air supply pressure to at least 70 psig.

b)Check back panel connec- tions. Raise sample supply to at least 5 psig but within ±2 psig of the span gas pressure.

5-2

TELEDYNE ANALYTICAL INSTRUMENTS

Page 56
Image 56
Teledyne 3160 manual Troubleshooting, Symptom Cause Correction