Appendix

Appendix

A.1 Specifications

(Subject to change without notice.)

Ranges: (Six ranges with AutoRanging on all ranges)

1:0–1 part per million (ppm) O2

2:0–10 ppm O2

3:0–100 ppm O2

4:0–1000 ppm O2

5:0–1 % O2

6:0–25 % O2

Accuracy: ±2% of full scale when calibrated and operated at the same constant temperature and pressure.*

±5% of full scale over the operating temperature range once tempera- ture equilibrium is reached.*

*Additional 0.04 ppm on 0-1 ppm range, generated by internal scrub- ber (which is necessary to maintain these specifications).

Response Time: 90 % of final reading in less than 61 seconds at 25 °C

Operating Temperature:0-50°C (32 -122 °F)

Analog Signal Outputs: For percent of full scale indication:

one 0–1 V dc non-isolated signal one 4–20 mA dc isolated signal

For range identification:

one 0–1 V dc non-isolated signal one 4–20 mA dc isolated signal

Digital Data Lines: One RS-232C serial interface for 2-way communications with separate host computer—for remote monitoring and control of all functions.

Alarms: Five user-programmable absolute-reading alarm setpoints, with Form C, SPDT relays (3A resistive).