Operations 4
When waiting for manual sensor settling during calibration zeroing, wait until the oxygen concentration has reached its lowest value and has remained so for
NOTE: 1. Zeroing in the
In scheduled zeroing and spanning (AutoZeroing and AutoSpanning), a time interval must be chosen. This interval determines how often the instru- ment zeros or spans itself. To change the scheduled time, edit it after setting the interval in the Set Up Clock Functions mode.
NOTE: 2. AutoZeroing and AutoSpanning may be delayed until the alarms have been enabled.
Calibration Using Span Gas: After zeroing, the instrument can now be calibrated using a span gas with a known oxygen concentration. The analyz- er may be calibrated on any range other than
NOTE: 3. Prior to calibration, allow the analyzer reading to come to a stable value with zero or sample gas flowing through the analyzer.
Select Active Sensor (Two cell blocks only): One or two O2 sensor blocks can be installed in the system. In dual sensor systems, either sensor can be isolated with pneumatic valves, and the other sensor chosen as the active sensor for O2 measurements. The system software is aware of how many sensor blocks are present, but will not automatically switch sensors should one fail.
This menu will not appear if only one cell block is installed.
Install Sensor and Test Alternate Sensor (Two cell blocks only): Pro- cedures are provided for installing a new sensor and for testing the alternate O2 sensor.
This menu will not appear if only one cell block is installed.
O2 Range
TELEDYNE ANALYTICAL INSTRUMENTS |