Communications Teledyne API T703/T703U Calibrator Operation Manual

122

STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK

addresses by pressing:

Internet Configuration Touchscreen Button Functions
BUTTON FUNCTION
[0] Press to cycle thr oug h the r ange of nu merals and
available characters ( 09” & . ”)
<CH CH> Moves the cursor one character left or right.
DEL Deletes a character at the cursor location.
ENTR Accepts the new setti ng and returns to the previous
me nu.
EXIT Ignores the new setting and returns to the previous
me nu.
Some buttons only appear when applicable.
SETUP X.X DHCP: OFF
SET> EDIT EXIT
SETUP X.X INST IP: 000.000.000.000
<SET SET> EDIT EXIT
SETUP X.X GATEWAY IP: 000.000.000.000
<SET SET> EDIT EXIT
SETUP X.X INST IP: [0] 00.000.000
<CH CH> DEL [0] ENTR EXIT
SETUP X.X SUBNET MASK:255.255.255.0
<SET SET> EDIT EXIT
SETUP X.X SUBNET MASK:[2]55.255.255.0
<CH CH> DEL [?] ENTR EXIT
SETUP X.X TCP PORT 3000
<SET EDIT EXIT
The PORT number needs to remain at 3000.
Do not change this setting unl ess i nstructe d to by
Teledyne Instruments Customer Service personnel.

From Step 1 above)

SETUP X.X GATEWAY IP: [0] 00.000.000
<CH CH> DEL [?] ENTR EXIT
Cursor
location is
indicated by
brackets
SETUP X.X INITIALIZING INET 0%
INITIALIZING INET 100%
SETUP X.X INITIALIZATI
O
N SUCCEEDED
SETUP X.X INITIALIZATION FAILED
SETUP X.X
COMM UN IC A T IONS ME NU
ID INET COM1 EXIT
Pressing EXIT from
any of the above
display menus
causes the Ethernet
to reinitialize its
internal interface
firmware

Contact your IT

Network Administrator