Teledyne API T703/T703U Calibrator Operation Manual Principles of Operation

197

9.2.3.2. Heater Control

The relay PCA controls the DC heaters for the O3 generator and photometer lamp

housing.

Figure 9-6: Heater Control Loop Block Diagram.

9.2.3.3. Relay PCA Status LEDs and Watch Dog Circuitry

Sixteen LEDs are located on the calibrator’s relay board (Figure 9-7) to indicate the

status of the calibrator’s heating zones and some of its valves; included is a general

operating watchdog indicator. Table 9-1 shows the states of these LEDs and their

respective functionality. Not all LEDs are used.

D15 (Green) - Photometer Lamp Heater
D16 (Green) – O3 Generator Lamp Heater
D7 (Green) Photometer Meas/Ref Valve
D9 (Green) – O3 Pump
D1 (RED)
Watchdog
Indicator
T703U only:
D10 (Green)
O3 Divert Valve
T703U only: D11 (Green) – O3 Generator Valve
D2
(
Yellow
)
Zero Air Pum
p
O
p
tion
D8 (Green) Zero Air Shutoff Valve Option

Figure 9-7: Status LED Locations – Relay PCA

07223C DCN6572