Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
93

4.12.3. LEVELS

A LEVEL is a combination of several parameters:
An ID number for the LEVEL
An action, (e.g. GENERATE, STANDBY)
A target concentration value
An output flow rate (if applicable)
Configuration for one or both of two status output blocks.
Up to twenty levels can be defined and used with the T703 using a range of ID numbers
from 0-98. Level 99 is reserved for standby. Are not time based and do not include
characteristics such as start time or duration, therefore a single LEVEL can not switch
between different concentration levels and flow rates. Separate flow and concentration
outputs must be programmed into separate LEVELs which are then individually started
and stopped either by an operator at the calibrator’s front panel or through a serial data
operation over the RS-232 or Ethernet ports.

4.12.4. ACTIVATING AN EXISTING LEVEL

To activate an existing defined LEVEL, press:
07223C DCN6572