APPENDIX A-3: Warnings and Test Functions Teledyne API - Models T703, T703U, 703E MENU TREES (05745E
DCN6572)

A-18

APPENDIX A-3: Warnings and Test Functions

Table A-2: Warning Messages

NAME 1 MESSAGE TEXT DESCRIPTION
WSYSRES SYSTEM RESET Instrument was power-cycled or the CPU was reset.
WDATAINIT DATA INITIALIZED Data storage was erased.
WCONFIGINIT CONFIG INITIALIZED Configuration storage was reset to factory configuration or
erased.
WPHOTOLTEMP PHOTO LAMP TEMP
WARNING Photometer lamp temperature outside of warning limits specified
by PHOTO_LAMP variable.
WO3GENTEMP O3 GEN LAMP TEMP
WARNING O3 generator lamp temperature outside of warning limits
specified by O3_GEN_LAMP variable.
WPHOTOREF PHOTO REFERENCE
WARNING Photometer reference reading less than 2500 mV or greater than
4999 mV.
WLAMPSTABIL PHOTO LAMP STABILITY
WARNING Photometer lamp reference step changes oc cur more than 25%
of the time.
WO3GENREF O3 GEN REFERENCE
WARNING O3 reference detector drops below 50 mV during reference
feedback O3 generator control.
WO3PUMP O3 PUMP W ARNING O3 pump failed to turn on within timeout perio d specified by
O3_PUMP_TIMEOUT variable.
WBOXTEMP BOX TEMP WARNING Chassis temperature outside of warning limits specified by
BOX_SET variable.
WREARBOARD REAR BOARD NOT DET Rear board was not detected duri ng power up.
WRELAYBOARD RELAY BOARD WARN Firmware is unable to communicate with the relay board.
WLAMPDRIVER LAMP DRIVER WARN Firmware is unable to communicate with either the O3 generator
or photometer lamp I2C driver chip.
WFRONTPANEL FRONT PANEL WARN Firmware is unable to communicate with the front pane l.
WANALOGCAL ANALOG CAL WARNING The A/D or at least one D/A channel has not been calibrated.
1 The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.