Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
57

To specify a starting time for the sequence, press:

STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the
calibrator is in standby
mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SEQUENCE CONFIGURATION
EDIT PRINT EXIT
SETUP X.X END OF SEQUENCES
INS PRNT EXIT
SETUP X.X 1) SEQ [NAME], [X] STEPS
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME:0
SET> EDIT EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown
in the message field
Scrolls back and forth between
existing sequences Continue pressing SET> until ...
This display only appears if there are no sequences currently
programmed into the calibrator.
OTHERWISE ...
SETUP X.X TIMER ENABLE:ENABLED
<SET SET> EDIT EXIT
SETUP X.X TIMER START: 01-JAN-06 00:00
<SET SET> EDIT EXIT
SETUP X.X TIMER START: 01-JAN-06 00:00
0 1 JAN 0 6 ENTR EXIT
EXIT discards the
new setting
ENTR accepts the
new setting
Toggle these
buttons to enter
starting day, month
and year.

DAY MONTH YEAR

SYSTEM TIME: 12:00
1 2 :0 0 ENTR EXIT

HOUR MINUTE

Toggle these
buttons to enter the
starting time
EXIT discards the
new setting
ENTR accepts the
new setting
07223C DCN6572