Teledyne API - Models T703, T703U, 703E MENU TREES (05745E DCN6572) APPENDIX A-2: Serial I/O Setup Variables

A-15

Setup Variable Numeric
Units Default
Value Value
Range Description
MODEM_INIT2 —
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0” 0
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM2 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
LINE_DELAY 1 ms. 0 0–1000
RS-232 inter-line transmit dela y
(0=disabled).
MACHINE_ID ID 0 0–9999
(Hessen: 0–
999)
Unique ID number for instrument.
COMMAND_PROMPT — “Cmd>
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes
(") when setting from the RS-232
interface.
TEST_CHAN_ID NONE,
O3 PHOTO
MEAS,
O3 PHOTO
REF,
O3 GEN REF,
OUTPUT
FLOW,
REGULATOR
PRESSURE,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
SAMPLE
TEMP,
PHOTO LAMP
TEMP,
O3 LAMP
TEMP,
CHASSIS
TEMP,
O3 PHOTO
CONC
Diagnostic analog output ID.
Enclose value in double quotes
(") when setting from the RS-232
interface.
PASS_ENABLE ON OFF, ON ON enables passwords.
OFF disables them.
DEF_CC_OUTPUT —
“00000000000
0” Any string of
exactly 12
characters
consisting of
the digits 0 and
1 only.
Default contact closure output
pattern when not executing a
sequence. Enclose value in
double quotes (") when setting
from the RS-232 interface.
PHOTO_LAMP_POWER mV 4500 0–5000 Photometer lamp power setting.
07223C DCN6572