Teledyne API - Models T703, T703U, 703E MENU TREES (05745E DCN6572) APPENDIX A-2: Serial I/O Setup Variables

A-13

Setup Variable Numeric
Units Default
Value Value
Range Description
O3_MIN_CONC PPB 25 0–100
O3 generator minimum reliable
concentration. Less than this is
treated as zero.
REF_DELAY Seconds 60 1–300
O3 generator reference feedback
control delay.
REF_FREQ Seconds 1 1–60
O3 generator reference
adjustment frequency.
REF_FSIZE Samples 4 1–10 O3 generator reference filter size.
REF_INTEG — 0.1 0–10
O3 generator reference PID
integral coefficient.
REF_DERIV — 0.2 0–10
O3 generator reference PID
derivative coefficient.
BENCH_DELAY Seconds 120 1–300
O3 generator bench feedback
control delay.
BENCH_FREQ Seconds 10 1–60
O3 generator bench adjustment
frequency.
BENCH_FSIZE Samples 3 1–10 O3 generator bench filter size.
BENCH_INTEG — 0.2 0–10
O3 generator bench PID integral
coefficient.
BENCH_DERIV — 0.5 0–10
O3 generator bench PID
derivative coefficient.
DRIVE_STABIL mV 10 0.1–100
O3 generator drive stability limit
to update concentration cache.
CACHE_RESOL PPB 2 0.1–20
O3 generator cache un-
normalized concentration
resolution.
O3_LAMP_CYCLE Seconds 2 0.5–30
O3 generator lamp temperature
control cycle period.
O3_LAMP_PROP 1/DegC 0.2 0–10
O3 generator lamp temperature
PID proportional coefficient.
O3_LAMP_INTEG Gain 0.01 0–10
O3 generator lamp temperature
PID integral coefficient.
O3_LAMP_DERIV Gain 0.2 0–10
O3 generator lamp temperature
PID derivative coefficient.
OUTPUT_FLOW_SLOPE — 1 0.001–100
Slope term to correct output flow
rate.
07223C DCN6572