Teledyne API T703/T703U Calibrator Operation Manual Principles of Operation
201

RELAY PCA

SENSOR SUITES
LOGIC DEVICES
(e.g. CPU, I2C bus,
Motherboard, etc.)
ON / OFF
SWITCH
PS 2
(+12 VDC)
Photometer
M/R valve
Cooling
Fan
PS 1
ANALOG SENSORS
O3Generator
Reference detector,
Photometer UV
Detector Pre-Amplifiers
& Amplifiers
Sensor Control
& I/O Logic
Solenoid
Drivers
KEY
AC POWER
DC POWER
AC
POWER IN
+5 VDC
±15 VDC
O3Generator UV
Lamp Xfromer
Photometer
UV Lamp P/S
GAS
PRESSURE
SENSORS
GAS
TEMPERATURE
SENSORS
O3Generator UV
Lamp P/S
O3Generator
UV Lamp
Photometer
Pump
Controlled
via I2C
DRY AIR
Pump
AC
Relay
Figure 9-8: Power Distribution Block diagram

9.2.6. AC POWER CONFIGURATION

The digital electronic systems will operate with any of the specified power regimes. As

long as instrument is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz

it will turn on and after about 30 seconds show a front panel display. Internally, the

status LEDs located on the Relay PCA, Motherboard and CPU should turn on as soon as

the power is supplied.

On the other hand, in some of the calibrators the dry air pump must be properly

configured for the type of power being supplied to the instrument. Figure 2-3 shows the

location of the Pump AC Configuration jumper.

07223C DCN6572