Teledyne API - Models T703, T703U, 703E MENU TREES (05745E DCN6572) APPENDIX A-2: Serial I/O Setup Variables

A-17

Setup Variable Numeric
Units Default
Value Value
Range Description
CLOCK_FORMAT —
“TIME=%H:%
M:%S” Any character
in the allowed
character set.
Up to 100
characters
long.
Time-of-day clock format flags.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
“%a” = Abbreviated weekday
name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format
(00 – 23).
“%I” = Hour in 12-hour format (01
– 12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal
number (01 – 12).
“%M” = Minute as decimal
number (00 – 59).
“%p” = A.M./P.M. indicator for
12-hour clock.
“%S” = Second as decimal
number (00 – 59).
“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as
decimal number.
“%%” = Percent sign.
FACTORY_OPT — 0 0–0x7fffffff
Factory option flags. Add values
to combine options.
1024 = enable software-
controlled maintenance mod e
2048 = enable Internet opti on 3
4096 = enable switch-controlled
maintenance mode
PWR_TIME Hours 0 0–500000 Total powered hours.
1 Dasibi emulation version onl y .
2 Experimental.
3 E Series internet option.
4 Hessen protocol versions onl y.
5 T703U only.

07223C DCN6572