Teledyne API T703/T703U Calibrator Operation Manual Specifications and Approvals

21

2. SPECIFICATIONS AND APPROVALS

2.1. SPECIFICATIONS

Table 2-1: T703/T703U System Specifications

PARAMETER SPECIFICATION

Linearity ± 1.0% of full scale
Precision 1.0 ppb
T703 T703U
Response Time
<180 seconds to 95% 240 seconds to 95%
Stability (7-days) <1% photometer feedback; <3% without photometer feedback (CNST or REF)
Operating Temperature Range 5-40ºC
Humidity Range 0 - 95% RH, non-condensing
Operating Altitude 10,000 ft Maximum
Dimensions (H x W x D) 7” (178 mm) x 17” (432 mm) x 24” (609 mm)
Weight 35.5 lbs (16.1 kg) including internal zero air pump
AC Power 100V – 120V, 60Hz, (64W); (with Zero Air Pump 138W)
220V – 240V, 50Hz, (81W); (with Zero Air Pump 157W)
Analog Output Ranges 0.1 V, 1 V, 5 V or 10 V (selectable)
Range with 5% under/over-range
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital control outputs
12 opto-isolated digital control inputs
8 opto-isolated digital status outputs
Optional I/O
1 USB com port
1 RS485
Multidrop RS232

Table 2-2: T703/T703U Specifications for Ozone Generator

T703 T703U
Flow Rate 1 to 5 LPM adjustable w/internal zero air source: 1 to 5 LPM adjustable
w/external zero air source: 1 to 15 LPM adjustable
Concentration Min: 50 ppb at 2 LPM
Max: 5 ppm at 1 LPM
Min: 3 ppb at 5 LPM
Max: 5 ppm at 1 LPM
Output Min: 100 ppb LPM
Max: 5 ppm LPM
Min: 15 ppb LPM
Max: 5 ppm LPM
Response Time: <180 sec. to 98% <240 sec. to 98%
Optical Feedback Standard
07223C DCN6572