Principles of Operation Teledyne API T703/T703U Calibrator Operation Manual
202
JP7
Pump
Configuration
Figure 9-9: Location of the AC Configuration Jumper for the Dry Air Pump

9.2.6.1. AC Configuration – Internal Pump (JP7)

AC power configuration for the internal dry air pump is set using Jumper set JP7.

Table 9-2: AC Power Configuration for Internal Pumps (JP7)
LINE
POWER
LINE
FREQUENCY
JUMPER
COLOR FUNCTION
JUMPER
BETWEEN
PINS
Connects pump pin 3 to 110 / 115 VAC power line 2 to 7
Connects pump pin 3 to 110 / 115 VAC power line 3 to 8
60 HZ WHITE
Connects pump pins 2 & 4 to Neutral 4 to 9
Connects pump pin 3 to 110 / 115 VAC power line 2 to 7
Connects pump pin 3 to 110 / 115 VAC power line 3 to 8
110VAC
115 VAC
50 HZ1 BLACK
Connects pump pins 2 & 4 to Neutral 4 to 9
Connects pump pins 3 and 4 together 1 to 6
60 HZ BROWN Connects pump pin 1 to 220 / 240VAC power line 3 to 8
Connects pump pins 3 and 4 together 1 to 6
220VAC
240 VAC 50 HZ1 BLUE Connects pump pin 1 to 220 / 240VAC power line 3 to 8
1 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the T300 and
has no function on the Models T700 or T703.
07223C DCN6572