Principles of Operation Teledyne API T703/T703U Calibrator Operation Manual

196

Power
Connections
for DC
Heaters
Status LED’s
(D2 through D16)
DC Power Supply
Test Points
Watchdog
Status LED (D1)
I2C Connector
DC
Valve Control
Drivers
AC Power
IN
DC Power
Distribution
Connectors
DC
V
alve &
Photometer
Pump Control
Connector
A
C Power
OUT to Dry
Air Pump
AC Pump
Configuration
Plug
Dry Air AC Pump
Control Relay
DC Valve &
Photometer
Pump
Control

Figure 9-5: Relay Board PCA with AC Relay Retainer Removed

This version of the Relay PCA includes one AC relay that controls the AC-powered Dry

Air (zero air) pump and A plastic insulating safety shield covers the remaining empty

AC Relay sockets.

CAUTION

NEVER REMOVE THIS SAFETY SHIELD WHILE THE INSTRUMENT IS PLUGGED IN AND

TURNED ON. THE CONTACTS OF THE AC RELAY SOCKETS BENEATH THE SHIELD

CARRY HIGH AC VOLTAGES EVEN WHEN NO RELAYS ARE PRESENT

9.2.3.1. Valve Control

The relay board also hosts two valve driver IC's, each of which can drive up four valves.

In the T703, the relay PCA controls only those valves associated with the O3 generator

and photometer.

In the T703U, the relay PCA controls two additional valves: the Divert valve and the O3

Gen valve.