Teledyne API - Models T703, T703U, 703E MENU TREES (05745E DCN 6572) APPENDIX A-6: MODBUS Registers

A-25

APPENDIX A-6: MODBUS Regist ers

MODBUS

Register

Address

(dec., 0-based)

Description Units

MODBUS Floating Point Input Registers

(32-bit IEEE 754 format; read in high-word, low-word order; read-only)

0 Output flow rate LPM
2 Photometer measured ozone concentration PPB
4 Ozone generator reference detector reading mV
6 Ozone generator lamp drive mV
8 Ozone generator lamp temperature °C
10 Regulator pressure PSIG
12 Internal box temperature °C
14 Photometer detector measure reading mV
16 Photometer detector reference reading mV
18 Photometer sample flow rate LPM
20 Photometer lamp temperature °C
22 Photometer sample pressure Inches Hg
24 Photometer sample temperature °C
26 Photometer slope computed during zero/spa n bench calibration
28 Photometer offset computed during zero/span bench calibration PPB
30 Ground reference mV
32 Precision 4.096 mV reference mV

MODBUS Discrete Input Registers

(single-bit; read-only)

0 System reset warning
1 Box temperature warning
2 Photometer lamp temperature warning
3 O3 generator lamp temperature warning
4 Photometer reference warning
5 Photometer lamp stability warning
6 O3 generator reference detector warning
7 Rear board communication warning
8 Relay board communication warning
9 O3 generator or photometer lamp I2C driver chip communication warning
10 Front panel communication warning
11 Analog calibration warning
12 System is OK (same meaning as SYSTEM_OK I/O signal)
13 O3 generator not yet stabilized
14 Calibration is active
15 Sequence is active
16 Regulator pressure warning