Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
102
current meter (see Figure 3-7 for pin assignments and diagram of the analog
output connector). This allows the use of a voltmeter connected across the
resistor to measure converter output as VDC or mVDC.
V
+DC Gnd
Figure 5-4: Alternative Setup Using 250 Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following
values:
Table 5-7: Current Loop Output Check
% FS Voltage across
Resistor for 2-20 mA
Voltage across
Resistor for 4-20 mA
0 500 mVDC 1000 mVDC