Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Maintenance Schedule & Procedures
187 07275A (DCN6005)
Table 10-1. T802 Maintenance Schedule
DATE PERFORMED ITEM ACTION FREQ CAL CHECK REQ’D.MANUAL
Particulate
Filter Replace Weekly or as
needed No 10.3.1
Verify Test
Functions
Record and
analyze
Weekly or after
any
Maintenance or
Repair
No 11.1.2
Pump
Diaphragm Replace Annually Yes 10.3.2
Perform Flow
Check Check Flow Annually No 10.3.4
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance or
Repair
No 10.3.3
Pneumatic
lines
Examine and
clean As needed Yes if
cleaned
Chassis Wipe down As needed
Only if
cover
removed
07275B DCN6418