Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
24
Verify that there is no apparent external shipping damage. If damage has
occurred, please advise the shipper first, then Teledyne API.
Included with your analyzer is a printed record of the final performance
characterization performed on your instrument at the factory.
This record, Final Test and Validation Data Sheet, PN 068350000, is an
important quality assurance and calibration record for this instrument. It should
be placed in the quality records file for this instrument.
1. Carefully remove the top cover of the analyzer and check for internal shipping
damage as follows:
a) Remove the locking screw located in the top, center of the Front panel;
b) Remove the two flat head, Phillips screws on the sides of the instrument
(one per side towards the rear);
c) Slide the cover backwards until it clears the analyzer’s front bezel;
d) Lift the cover straight up.
2. Inspect the interior of the instrument to make sure all circuit boards and other
components are in good shape and properly seated.
3. Check that the connectors of the various internal wiring harnesses and
pneumatic hoses are firmly and properly seated.
4. Verify that all of the optional hardware ordered with the unit has been installed.
These are listed on the paperwork accompanying the analyzer.