Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
106
SAMPLE RANGE=100.00 % O2=XXX.XX
<TST TST> CAL SETUP
DIAG TEST CHAN:NONE
PREV NEXT ENTR EXIT
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG TEST CHAN OUTPUT
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
Toggle these buttons
to choose a mass flow
controller TEST
channel parameter DIAG TEST CHANNEL:SAMPLE PRESSURE
PREV NEXT ENTR EXIT EXIT discards the new
settingENTR accepts the new setting
SETUP X.X PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTR EXIT
Toggle these
buttons to enter
the correct
PASSWORD
(The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory).