Introduction, Features and Options Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 20 Option Option Number Description/Notes Reference
USB COM Port
64A
Separate option if instrument not configured with Option 64B (analog
inputs). Disabled when using Multidrop or RS-485 communication.
Sections 3.3.1.8 and
6.4
Second Gas Sensor
67A Carbon Dioxide (CO2) Sensor 0-20% Sections 9.5, 12.2,
12.2.1 and 12.3.1
Special Features Built in features, software activated
N/A
Maintenance Mode Switch, located inside the instrument, places the
analyzer in maintenance mode where it can continue sampling, yet
ignore calibration, diagnostic, and reset instrument commands. This
feature is of particular use for instruments connected to Multidrop or
Hessen protocol networks.
Call Technical Support for activation.
N/A
N/A
Second Language Switch activates an alternate set of display
messages in a language other than the instrument’s default language.
Call Technical Support for a specially programmed Disk on Module containing
the second language.
N/A
N/A
Dilution Ratio Option allows the user to compensate for diluted
sample gas, such as in continuous emission monitoring (CEM) where
the quality of gas in a smoke stack is being tested and the sampling
method used to remove the gas from the stack dilutes the gas.
Call Technical Support for activation.
Sections 3.4.3.2
and 5.4.4