Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
180 07275A (DCN6005)
9.5.3. ACTIVATE CO2 SENSOR STABILITY FUNCTION

To change the stability test function from O2 concentration to the CO2 sensor

output, press:

Press 3

times to return

to

menu

SETUP X.X
CO2 O2 ENTR
<TST TST> SETUP
SETUP X.X
CFG DAS ACAL RANG PASS CLK EXIT
SETUP X.X
COMM DIAG ALRM EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV JUMP EDIT PRNT EXIT
SETUP X.X
<PREV NEXT> JUMP PRNT EXIT
SETUP X.X
O2 ENTR EXIT
SETUP X.X
EXIT
CO2and O2
options only
appear if
associated
sensors are
installed.
NOTE
Use the same procedure to reset the STB test function to O2 when the CO2 calibration procedure is
complete.