8.3. REMOTE ACCESS BY MODEM
The T802 can be connected to a modem for remote access. This requires a cable
between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M
cable (available from Teledyne API with PN WR0000024).
Once the cable has been connected, check to make sure:
The DTE-DCE switch on the rear panel is in the DCE position.
The T802 COM port is set for a baud rate that is compatible with the modem,
The Modem is designed to operate with an 8-bit word length with one stop bit.
The MODEM ENABLE communication mode is turned on (Mode 64, see
Section 6.2.1).
Once this is completed, the appropriate setup command line for your modem can
be entered into the analyzer. The default setting for this feature is
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0
This string can be altered to match your modem’s initialization and can be up to
100 characters long.
To change this setting press:
(The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory).