Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
218
QUESTION ANSWER
How do I perform a leak check? See Section 10.3.3.
How do I measure the sample
flow?
Sample flow is measured by attaching a calibrated rotameter, wet
test meter, or other flow-measuring device to the sample inlet port
when the instrument is operating. The sample flow should be 120
cm3/min 10%. See Section 9.4.2.
What is the averaging time for
this instrument?
The default averaging time, optimized for ambient pollution
monitoring, is 60 seconds for stable concentrations and 10 seconds
for rapidly changing concentrations; See 12.5.9 for more information.
11.9. TECHNICAL ASSISTANCE

If this manual and its troubleshooting / service sections do not solve your

problems, technical assistance may be obtained from: