Appendix C Warranty/Repair Questionnaire
T80X, M80XE
(06532C DCN 5798)
TELEDYNE API CUSTOMER SERVICE
EMAIL: api-customerservice@teledyne.com
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER: _______________________________ PHONE: _____________________________________
CONTACT NAME: ___________________________ FAX NO. _____________________________________
SITE ADDRESS: ___________________________________________________________________________
MODEL TYPE: ______________ SERIAL NO.: ________________ FIRMWARE REVISION: ____________
Are there any failure messages? _______________________________________________________________
________________________________________________________________________________________________________________________________
________________________________________________________________________ (Continue on back if necessary)
PLEASE COMPLETE THE FOLLOWING TABLE:
PARAMETER RECORDED VALUE ACCEPTABLE VALUE
O2 RANGE1 % 0-100%
O2 CELL TEMP1 ºC 50 ± 5
O2 SLOPE1 1.0 ± 0.3
O2 OFFSET1 -10 to 10%
CO2 RANGE1 % 0 to 20%
CO2 CELL TEMP1 ºC 50 ± 5
CO2 SLOPE1 1.0 ± 0.3
CO2 OFFSET1 -10 to 10%
STABIL % 0.2% with zero air
PRESS in-Hg-A ambient ± 1
SAMPLE FLOW cm3/min 120 ± 20
BOX TEMP ºC ambient ± 5ºC
following values are under the signal i/o submenu
REF_4096_MV mV 4096mV ±2 mV and Must be Stable
REF_GND mV 0± 0.5 and Must be Stable
1 Not all models are equipped with both an O2 and a CO2 sensor.
Cap the SAMPLE inlet and record the flow rate and pressure readings:
What is PRESS____________________in-Hg-A
What is the SAMPLE FLOW__________ cc/min
What are the failure symptoms? __________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
What test have you done trying to solve the problem? ________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
07275B DCN6418