Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual
144
7.1.4. SETUP DAS EDIT: ACCESSING THE DAS EDIT MODE
DAS configuration is most conveniently done through the APICOM remote
control program. The following list of key strokes shows how to edit using the
front panel.
SETUP X.X
CFG ACAL RNGE PASS CLK MORE EXIT
SETUP X.X
VIEW EXIT
SETUP X.X
NEXT INS DEL PRNT
SETUP X.X
8 1 8 EXIT
Enters EDIT mode for the selected channel
<TST TST> CAL MSG
Exports the configuration of all data channels to the
RS-232 interface
Selects the previous data channel in the list
Inserts a new data channel into the list BEFORE the
selected channel
Selects the next data channel in the list
Deletes the currently selected data channel
Enters EDIT mode
When editing the data channels, the top line of the display indicates some of the
configuration parameters.
For example, the display line:
0) CONC: ATIMER, 2, 4032, RS232
Translates to the following configuration:
0 Channel No.
CONC Channel Name
ATIMER Trigger Event
2 Parameters – number of parameters included in this channel
4032 Event – number of data points this channel is set up to store
RS-232 Port via which values automatically reported