Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual

86

Table 5-2: Variable Names (VARS)

NO. VARIABLE DESCRIPTION ALLOWED

SETTINGS

VARS
Default
settings
0 DAS_HOLD_OFF
Changes the internal Data Acquisition System (DAS) HOLD
OFF timer:
No data are stored in the DAS channels during situations
when the software considers the data to be questionable
such as during warm up or just after the instrument returns
from one of its calibration modes to SAMPLE Mode.
May be set for
intervals
between
0.5 – 20 min

15 min.

1 STABIL_GAS1 Selects which gas measurement is displayed when the
STABIL test function is selected. O2; CO2 O2
2 TPC_ENABLE
NOTE: It is strongly recommended that this variable
NOT be changed.
ON enables, OFF disables temperature and pressure
compensation
ON, OFF ON
3 DYN_ZERO
Special configuration;
consult factory
[Automatically adjusts offset and slope of the O2 response
when performing a zero point calibration during an AutoCal.] [ON, OFF] [OFF]
4 DYN_SPAN
Special configuration;
consult factory
[Automatically adjusts offset and slope of the O2 response
when performing a zero point calibration during an AutoCal.] [ON, OFF] [OFF]
5 CONC_PRECISION
Allows the user to set the number of significant digits to the
right of the decimal point display of concentration and
stability values.
AUTO, 1, 2,
3, 4 AUTO
6 CLOCK_ADJ Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.
-60 to +60
s/day 0 sec
7 SERVICE_CLEAR
Pressing the OFF button to display SERVICE_CLEAR:ON,
followed by pressing ENTR resets the service interval timer
and returns this Var back to its default setting, ready for the
next reset.
ON, OFF OFF
8 TIME_SINCE_SVC Displays time in hours since last service (restarted by the
SERVICE_CLEAR Variable). 0-50,000 0
9 SVC_INTERVAL Sets the interval in hours between service reminders. 0-100,000 0
1 This VAR only appears if the optional CO2 sensor is installed.