Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A

Error! Unknown document property name.Error! Unknown document property name. A-13

Setup Variable Numeric

Units Default

Value Value Range Description

MODEM_INIT —
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM1 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_MODE2 BitFlag 0 0–65535 RS-232 COM2 mode flags.
(Same settings as
RS232_MODE.)
BAUD_RATE2 — 19200 300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM2 baud rate.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
MODEM_INIT2 —
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM2 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
MACHINE_ID ID 802 0–9999 Unique ID number for instrument.
COMMAND_PROMPT — “Cmd>
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
TEST_CHAN_ID — NONE
NONE, SAMPLE PRESSURE, SAMPLE FLOW, O2 CELL TEMP 3, CO2 CELL TEMP 1, CHASSIS TEMP
Diagnostic analog output ID.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
07275B DCN6418